Results From a Second-Generation Vestibular Implant in Human Subjects: Diagnosis May Impact Electrical Sensitivity of Vestibular Afferents

Conclusion: The etiology of vestibular loss appears to have a profound impact on sensitivity of vestibular afferents in distinction to cochlear afferents. If this dichotomy is common, it may limit the application of vestibular implants to diagnoses with preserved sensitivity of vestibular afferents. We speculate it is due to differences in topographic organization of Scarpa's versus the spiral ganglion. In two subjects, the second-generation device can produce higher velocity eye movements than seen in the four subjects receiving the first-generation device.