ConclusionCorrect analysis of the acquired ECG signal is the main challenge for cardiologist due to involvement of various types of noises (high and low frequency). Twenty two real time ECG records have been evaluated based on various FoM such as SNR, Se, Acc and DER for the proposed FrWT and existing FrFT and IPCA preprocessing techniques. Acquired real-time ECG database in normal and disease situations is used for the purpose. The values of FoMs indicate high SNR and better detection of R-peaks in a ECG signal which is important for the diagnosis of cardiovascular disease. The proposed FrWT outperforms all other techniques and holds both analytical attributes of the actual ECG signal and alterations in the amplitudes of various ECG waveforms adequately. It also provides signal portrayals in the time-fractional-frequency plane with low computational complexity enabling their use practically for versatile applications.Graphical abstract