Quantitative outcome of registration methods for correcting cardiac drift in cardiac PET/CT imaging

This study evaluates three postimaging registration techniques to correct PET/CT misalignment by altering the transmission map to match myo­cardial uptake. Simulated misalignment studies were performed with a cardiac torso phantom filled with [18F]FDG at 10:1 myocardium/background. An air-filled saline bag affixed to the medial left lung surface served as a distensible lung. An initial CT acquisition was followed by successive PET acquisitions consisting of small displacements of the cardiac insert into the left lung. Phantom transmission scans were aligned to the myocardial uptake in the emission scans by applying 1) full rigid-body translations and rotations, 2) rigid-body restricted to medial / lateral and superior / inferior translation, or 3) an emission-driven method that adds myocardial tissue to the transmission scan. These methods were also applied to 10 low-likelihood coronary artery disease (CAD) patients showing signs of cardiac drift. Full rigid-body registration showed significant over-correction (p < 0.004) of activity concentrations in the artifact areas of the phantom data due the relocation of highly attenuating structures (i.e., spine). Inaccurate regional activity distributions were also observed as streaks extending from the spine and these results were replicated in the patient population. There was no significant difference between the true phantom activity concentration after correction with the emission-driven method. Misalignment corrected with t...