Collapse of zone 0 landing TEVAR (Najuta) and the development of higher brain dysfunction

J Cardiovasc Surg (Torino). 2024 Apr 15. doi: 10.23736/S0021-9509.24.12982-5. Online ahead of print.ABSTRACTAlthough thoracic endovascular repair (TEVAR)-specific complications often develop, stent-graft collapse is a rare, but fatal complication that requires attention. A 62-year-old male underwent TEVAR for a saccular distal arch aortic aneurysm. After the placement of the Gore TAG (W. L. Gore & Associates, Inc., Newark, DE, USA) from zones 2 to 4, a Najuta endograft (Kawasumi Laboratories, Inc., Tokyo, Japan) was deployed from zone 0. Neither intraoperative angiography nor postoperative contrast-enhanced computed tomography (CT) showed endoleaks or migration. On day 12 after surgery, the patient suddenly lost consciousness during rehabilitation, and CT revealed the collapse of the Najuta endograft. In emergency surgery, the Najuta endograft was removed and the TAG was sutured to the aorta. Although the patient survived, he developed irreversible higher brain dysfunction. The cause of the collapse was examined by the manufacturer and only a slight bird-beak configuration was noted. There were no other findings to indicate the cause of the collapse. The Najuta endograft is a semi-customized system that is created according to the three-dimensional morphology of each individual aortic arch and, thus, is expected to follow the flexion of this vessel. Nevertheless, the risk of collapse needs to be considered.PMID:38618698 | DOI:10.23736/S0021-9509.24.12982-5
Source: The Journal of Cardiovascular Surgery - Category: Cardiovascular & Thoracic Surgery Authors: Source Type: research