Clinical validation of electrochemical biosensor for the detection of methylglyoxal in subjects with type-2 diabetes mellitus

This study was designed to determine the utility of the developed electrochemical biosensor to measure the level of MG in human plasma from type-2 diabetes mellitus patients. Electrochemical studies were carried out with optimized experimental parameters using the modified Platinum-electrode. Subsequently, clinical studies using 350 blood plasma samples were conducted and the results were validated against the ELISA kit, Normal Glucose Tolerance (NGT), and glycosylated haemoglobin (HbA1c). The MG sensor exhibited a linear range of 1.0-7.5 μM concentration with a sensitivity of 1.02 mA µM-1, a limit of detection of 0.21 µM, a limit of quantification of 0.70 µM and a response time less than 10 s. The sensor showed 90% correlation with ELISA data. The developed biosensor showed a significant correlation with HbA1c and fasting plasma glucose suggesting that it can be used as a point-of-care device to screen for diabetes.PMID:37951008 | DOI:10.1016/j.bioelechem.2023.108601