Application of probe electrospray ionization-tandem mass spectrometry to ultra-rapid determination of glufosinate and glyphosate in human serum

In this study, we addressed this limiting factor by developing an ultra-rapid and simple analytical technique using a combination of probe electrospray ionization (PESI) and tandem mass spectrometry (MS/MS), which gives quantitative results within 0.3 min. Herbicide standards were added to human serum that was then subjected to analysis (N = 5 per concentration). The analysis was repeated daily over eight consecutive days. The limit of detection (LOD) was 0.59 μg/mL for glufosinate and 0.20 μg/mL for glyphosate. The limit of quantitation (LOQ), i.e., the lowest point on the calibration curves, was 1.56 μg/mL for both the herbicides. The matrix effects were observed at three different concentrations (between 95.7%–104% for glufosinate, and between 90.7%–95.7% for glyphosate). When applied to samples taken from actual poisoning cases (six samples for each herbicide), the present method gave almost the same quantitative values as those obtained by conventional high-performance liquid chromatography with fluorescence detection. Thus, we believe that PESI-MS/MS could emerge as a rapid diagnosis method in the clinical emergency field.
Source: Journal of Pharmaceutical and Biomedical Analysis - Category: Drugs & Pharmacology Source Type: research