NEMA NU-4 performance evaluation of a non-human primate animal PET.

This article presents a performance evaluation of the PET scanner according to the National Electrical Manufacturers Association NU-4 2008 standards. All measurements were made for an energy window of 360~660 keV and a coincidence timing window of 2 ns. In terms of the FWHM, the FBP reconstructed spatial resolution results in all three directions at the radial position of 5 mm were better than or approached to 2 mm, and remained below 3.0 mm within the central 5 cm diameter of the FOV. The peak absolute sensitivity of the scanner was measured 1.80%. For a monkey-sized phantom, the scatter fraction was 34.2% and the peak noise equivalent count rate (NECR) was 26.5 kcps at 64.3 kBq/cc. The overall imaging capabilities of the scanner were also assessed using in vivo imaging study of a rhesus macaque. The performance measurements demonstrate that the Eplus-260 primate PET scanner has the potential ability to obtain good quality and high-contrast images for non-human primates, especially the brain of large non-human primates and could be considered as one technologically advanced dedicated non-human primate PET scanner available today. PMID: 30947155 [PubMed - as supplied by publisher]