MRI Compatible Lumbopelvic Movement Measurement System to Validate and Capture Task Performance During Neuroimaging

This study examined the concurrent validity of air pressure changes between the digital and analog sensors. The digital and analog data were compared in 23 participants during the performance of modified bilateral and unilateral right and left hip bridges. Spearman’s correlations were calculated for each sensor during the three bridging tasks and showed high positive correlations, indicating that over 87% of pressure change from the analog gauge can be explained by the pressure from the digital sensor. Bland-Altman plots showed no bias and mean differences were under three mmHg. This pressure system improves the rigor of future studies by validating the digital data from the system and increasing the capabilities of capturing lumbopelvic task performance occurring inside the scanner bore.
Source: IEE Transactions on Neural Systems and Rehabilitation Engineering - Category: Neuroscience Source Type: research