Commissioning of a 3D image-based treatment planning system for high-dose-rate brachytherapy of cervical cancer

The objective of this work is to present commissioning procedures to clinically implement a three-dimensional (3D), image-based, treatment-planning system (TPS) for high-dose-rate (HDR) brachytherapy (BT) for gynecological (GYN) cancer. The physical dimensions of the GYN applicators and their values in the virtual applicator library were varied by 0.4 mm of their nominal values. Reconstruction uncertainties of the titanium tandem and ovoids (T&O) were less than 0.4 mm on CT phantom studies and on average between 0.8–1.0 mm on MRI when compared with X-rays. In-house software, HDRCalculator, was developed to check HDR plan parameters such as independently verifying active tandem or cylinder probe length and ovoid or cylinder size, source calibration and treatment date, and differences between average Point A dose and prescription dose. Dose-volume histograms were validated using another independent TPS. Comprehensive procedures to commission volume optimization algorithms and process in 3D image-based planning were presented. For the difference between line and volume optimizations, the average absolute differences as a percentage were 1.4% for total reference air KERMA (TRAK) and 1.1% for Point A dose. Volume optimization consistency tests between versions resulted in average absolute differences in 0.2% for TRAK and 0.9 s (0.2%) for total treatment time. The data revealed that the optimizer should run for at least 1 min in order to avoid more than 0.6% dwell time change...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Source Type: research