A comparative analysis of deep learning-based location-adaptive threshold method software against other commercially available software

This study aimed to analyze the performance of the coronary artery segmentation of a software platform with a deep learning-based location-adaptive threshold method (DL-LATM) against commercially available software platforms using CCTA. The dataset from intravascular ultrasound (IVUS) of 26 vessel segments from 19 patients was used as the gold standard to evaluate the performance of each software platform. Statistical analyses (Pearson correlation coefficient [PCC], intraclass correlation coefficient [ICC], and Bland-Altman plot) were conducted for the lumen or plaque parameters by comparing the dataset of each software platform with IVUS. The software platform with DL-LATM showed the bias closest to zero for detecting lumen volume (mean difference = -9.1 mm3, 95% confidence interval [CI] = -18.6 to 0.4 mm3) or area (mean difference = -0.72 mm2, 95% CI = -0.80 to -0.64 mm2) with the highest PCC and ICC. Moreover, lumen or plaque area in the stenotic region was analyzed. The software platform with DL-LATM showed the bias closest to zero for detecting lumen (mean difference = -0.07 mm2, 95% CI = -0.16 to 0.02 mm2) or plaque area (mean difference  = 1.70 mm2, 95% CI  = 1.37 to 2.03 mm2) in the stenotic region with significantly higher correlation coefficient than other commercially available software platforms (p <  0.001). The result shows that the software platform with DL-LATM has the potential to serve as an aiding system for CAD evaluation.
Source: The International Journal of Cardiovascular Imaging - Category: Radiology Source Type: research