In Vitro Comparison of a Closed and Semi-closed Leaflet Design for Adult and Pediatric Transcatheter Heart Valves

Ann Biomed Eng. 2024 Apr 13. doi: 10.1007/s10439-024-03502-3. Online ahead of print.ABSTRACTTranscatheter heart valve replacements (TVR) are mostly designed in a closed position (c) with leaflets coaptating. However, recent literature suggests fabricating valves in semi-closed (sc) position to minimize pinwheeling. With about 100,000 children in need of a new pulmonary valve each year worldwide, this study evaluates both geometrical approaches in adult as well as pediatric size and condition. Three valves of each geometry were fabricated in adult (30 mm) and pediatric (15 mm) size, using porcine pericardium. To evaluate performance, the mean transvalvular pressure gradient (TPG), effective orifice area (EOA), and regurgitation fraction (RF) were determined in three different annulus geometries (circular, elliptic, and tilted). For both adult-sized valve geometries, the TPG (TPGC = 2.326 ± 0.115 mmHg; TPGSC = 1.848 ± 0.175 mmHg)* and EOA (EOAC = 3.69 ± 0.255 cm2; EOASC = 3.565 ± 0.025 cm2)* showed no significant difference. Yet the RF as well as its fluctuation was significantly higher for valves with the closed geometry (RFC = 12.657 ± 7.669 %; RFSC = 8.72 ± 0.977 %)*. Recordings showed that the increased backflow was caused by pinwheeling due to a surplus of tissue material. Hydrodynamic testing of pediatric TVRs verified the semi-closed geometry being favourable. Despite the RF (RFC = 7.721 ± 0.348 cm2; RFSC = 5.172 ± 0.679 cm2), these valves also showed an improved...