Inter-laboratory comparison of routine autoantibody detection methods for autoimmune neuropathies and myasthenia gravis

AbstractSerological tests are important to detect autoantibodies (autoAbs) in patients with autoimmune neuropathies (AN) and myasthenia gravis (MG) as they are biomarkers for diagnosis, stratification, treatment selection, and monitoring. However, tests to detect autoAbs frequently lack proper standardization and results differ across diagnostic laboratories. We compared results for tests routinely performed in Spanish diagnostic laboratories to detect AN and MG autoAbs. In the Spanish Society of Immunology Autoimmunity Group national workshop, serum samples from 13 patients with AN or MG were tested for anti-ganglioside, anti-myelin-associated glycoprotein (MAG), anti-nicotinic acetylcholine receptor (AChR), and anti-muscle-specific kinase (MuSK) autoAbs using reference methods and were distributed for analysis to 27 participating laboratories using their routine methods. Overserved were inter-laboratory variability and worryingly low sensitivity, especially for anti-ganglioside immunoglobulin G and anti-MAG autoAb detection. This pilot study reflects autoAbs detection state of the art in AN and MG testing in leading diagnostic laboratories in Spain, highlighting the need for standardization prior to clinical use.
Source: Journal of Neurology - Category: Neurology Source Type: research