X-ray Doses in Relation to Body Mass, Indication, and Substrate During Pediatric Electrophysiological Procedures on the Heart

AbstractThe main goal of this study is to determine typical values of dose area product (DAP) and difference in the effective dose (ED) for pediatric electrophysiological procedures on the heart in relation to patient body mass. This paper also shows DAP and ED in relation to the indication, the arrhythmia substrate determined during the procedure, and in relation to the reason for using radiation. Organ doses are described as well. The subjects were children who have had an electrophysiological study done with a 3D mapping system and X-rays in two healthcare institutions. Children with congenital heart defects were excluded. There were 347 children included. Significant difference was noted between mass groups, while heavier children had higher values of DAP and ED. Median DAP in different mass groups was between 4.00 (IQR 1.00 –14.00) to 26.33 (IQR 8.77–140.84) cGycm2. ED median was between 23.18 (IQR 5.21 –67.70) to 60.96 (IQR 20.64–394.04) µSv. The highest DAP and ED in relation to indication were noted for premature ventricular contractions and ventricular tachycardia—27.65 (IQR 12.91–75.0) cGycm2 and 100.73 (IQR 53.31 –258.10) µSv, respectively. In arrhythmia substrate groups, results were similar, and the highest doses were in ventricular substrates with DAP 29.62 (IQR 13.81–76.0) cGycm2 and ED 103.15 (IQR 60.78 –266.99) µSv. Pediatric electrophysiology can be done with very low doses of X-rays when using 3D mapping systems compared to X-rays-based...
Source: Pediatric Cardiology - Category: Cardiology Source Type: research