Evaluation of attenuation correction method for head holder in brain perfusion single-photon emission computed tomography

AbstractHead holder attenuation affects brain perfusion single-photon emission computed tomography (SPECT) image quality. Here, we proposed a head holder-attenuation correction (AC) method using attenuation coefficient maps calculated by Chang ’s method from CT images. Then, we evaluated the effectiveness of the head holder-AC method by numerical phantom and clinical cerebral perfusion SPECT studies. In the numerical phantom, the posterior counts were 10.7% lower than the anterior counts without head holder-AC method. However, by perfor ming head holder-AC, the posterior count recovered by approximately 6.8%, approaching the true value. In the clinical study, the normalized count ratio was significantly increased by performing the head holder-AC method in the posterior-middle cerebral artery, posterior cerebral artery and cerebellu m regions. There were no significant increases in other regions. The head holder-AC method can correct the counts attenuated by the head holder.
Source: Radiological Physics and Technology - Category: Physics Source Type: research