True volumetric counting of CD34+ cells using flow cytometry

In this study we have used a direct true volumetric counting of CD34+ cells under continuous flow pump to overcome potential drawbacks with impact in rare cell analysis. To confirm this hypothesis, we have compared the results of CD34+ cell enumeration using non-volumetric vs. volumetric systems with FC500 (Beckman Coulter) and Attune NxT (ThermoFisher) flow cytometers, respectively, in mobilized peripheral blood samples. No statistically significant differences were observed between measurements of CD34+ cells using beads, when the FC500 and Attune NxT absolute counting values were compared, or when CD34+ counts were compared on the Attune NxT, either using or not using beads. Linear regressions to study the relationship between volumetric and non-volumetric CD34+ counts confirmed the accuracy of each method. Bland-Altman test showed agreement between both methods. Our data showed that CD34+ cell enumeration using a volumetric system is comparable with current counting systems. This method represents an alternative with the advantage of the simplification of sample preparation and the reduction of the analysis subjectivity.PMID:38395105 | DOI:10.1016/j.jim.2024.113649