Setup error of electronic portal image device in IMRT for thoracic tumors and its influence

CONCLUSIONS: In general, the first treatment of radiotherapy in thoracic tumors mostly has a certain degree of setup error, which is most evident in the X direction. When the CTV → PTV outward expansion distance is set at 7, 8, and 7 mm in the X direction, Y direction, and Z direction, respectively, it can effectively ensure that 95% ITV reach the prescribed dose in 90% of patients in the presence of a setup error. EPID helps to achieve the desired effect of radiotherapy, improves the efficacy of radiotherapy, and reduces the side effects caused by radiotherapy errors.PMID:38164863 | DOI:10.26355/eurrev_202312_34799