Optimization of 4D Flow MRI Spatial and Temporal Resolution for Examining Complex Hemodynamics in the Carotid Artery Bifurcation
ConclusionA spatial resolution of 0.74 –1.00 mm3 and a temporal resolution of 23 –48 ms (1–2 k-space segments) provides a 4D flow MRI protocol capable of imaging velocity and TAWSS in regions of complex flow within the carotid bifurcation at a clinically acceptable scan time.
Source: Cardiovascular Engineering and Technology - Category: Cardiology Source Type: research
More News: Angiography | Cardiology | Cardiovascular | CT Scan | Heart | PET Scan | Statistics | Stroke