IJERPH, Vol. 19, Pages 3548: A Simple Method to Quantify Outward Leakage of Medical Face Masks and Barrier Face Coverings: Implication for the Overall Filtration Efficiency

This study reports an original method to quantify outward leakage and how wearing style impacts on leaks and filtration efficiency. The amount of air leakage was evaluated on four medical masks and four barrier face coverings, exploiting a theoretical model and an instrumented dummy head in a range of airflows between 30 and 160 L/min. The fraction of air leaking at the face seal of the medical masks and barrier face coverings ranged from 43% to 95% of exhaled air at 30 L/min and reduced to 10–85% at 160 L/min. Filter breathability was the main driver affecting both leak fraction and total filtration efficiency that varied from 5% to 53% and from 15% to 84% at 30 and 160 L/min, respectively. Minor changes were related to wearing style, supporting indications on the correct mask use. The fraction of air leaking from medical masks and barrier face coverings during exhalation is relevant and varies according to design and wearing style. The use of highly breathable filter materials reduces air leaks and improve total filtration efficiency.
Source: International Journal of Environmental Research and Public Health - Category: Environmental Health Authors: Tags: Article Source Type: research