Left atrial assist device for heart failure with preserved ejection fraction: initial results with torque control mode in diastolic heart failure model

In this study, we tested four different TC settings (TC modes 0.9, 1.0, 1.25, and 1.5) using an in vitro mock circulatory loop. Mild, moderate, and severe diastolic heart failure (DHF) conditions, as well as normal heart condition, were simulated with the four TC modes. Also, we evaluated the LAAD in vivo with three calves. The LAAD was implanted at the mitral position with four TC settings (TC modes 0.9, 1.0, 1.1, 1.2). With LAAD support, the in vitro cardiac output and aortic pressure recovered to normal heart levels at TC 1.25 and 1.5 even under severe DHF conditions with little pump regurgitation. The TC mode tested in vivo with three calves, and it also showed favorable result without elevating the left ventricular end-diastolic pressure. These initial in vitro and in vivo results suggest that the TC mode could be potentially effective, and the LAAD could be a treatment option for HFpEF patients.