Three-Dimensional Planning of the Mandibular Margin in Hemifacial Microsomia Using a Printed Patient-Specific Implant

Hemifacial microsomia (HFM) is also known as malformation of the first and second branchial arches (BAs), oculoauriculovertebral dysplasia syndrome, and Goldenhar syndrome. Possibilities resulting from CAD/CAM may be a valuable tool to existing procedures to treat these abnormalities. The aim of this study was to report a case where the planning and implementation of an intraorally inserted 2-part patient-specific mandible implant was used for the treatment of HFM esthetic remained problem after orthognathic surgery. A 20-year-old female patient who continued to suffer from the esthetic defect of the unilateral hypoplastic mandible after completion of the orthognathic surgery attended for consultation. Using CT scan and the software Geomagic Freeform (3D Systems, USA), a 2-piece titanium implant was designed and printed to restore the osseous frame of the basal border of the mandible. The base was made of solid polished titanium to minimize soft tissue abrasion. Due to its split design, the implant could be placed anatomically exactly at the mandibular margin via an intraoral access and to avoid damage of the mental nerve. There were no postoperative complications such as infections, soft tissue reactions to the implant, sensitivity disorders, or dehiscence. The occlusion was regular. A measurement of the postoperative x-rays showed a clear increase in bony symmetry. No postoperative pain or trismus was seen. Functionally, the therapy ended with the completion of wound healin...
Source: Journal of Craniofacial Surgery - Category: Surgery Tags: Clinical Studies Source Type: research