Cut-off value of mal-apposition volume and depth for resolution at early phase of acute incomplete stent apposition after CoCr-EES implantation

AbstractThe purpose of this study was to clarify a cut-off value for acute incomplete stent apposition (ISA) volume and maximum-depth to predict ISA resolution at 1- and 3-month follow-up in patients treated with cobalt –chromium everolimus-eluting stents. In total, 95 cases and 103 stents were registered in the MECHANISM-Elective sub-study. Acute ISA-volume was measured by the trapezoid rule. ISA resolution of cut-off value at 1- and 3-month was estimated by ISA-volume and maximum-depth using receiver operatorat ing characteristic curve analysis. The total number of analysed acute ISAs was 202 in the 1-month group and 225 in the 3-month group. A total of 123 ISAs at 1-month and a total of 169 ISAs at 3-month had been resolved. The cut-off value of ISA resolution by ISA-volume was 0.169 mm3 at 1-month (AUC: 0.725, sensitivity: 72.2%, specificity: 61.0%) and 0.295 mm3 at 3-month (AUC: 0.757, sensitivity: 75.0%, specificity: 60.4%). The cut-off value of ISA resolution by ISA maximum-depth demonstrated was 0.285  mm at 1-month (area under curve (AUC): 0.789, sensitivity: 70.9%, specificity: 69.9%) and 0.305 mm at 3-month (AUC: 0.663, sensitivity: 60.7%, specificity: 66.9%). Incidence of ISA resolution was significantly lower in combination with cut-off values of ISA-volume and maximum-depth (33%,p <  0.001, at 1-month; 56%,p = 0.003, at 3-month). Combining the cut-off value of ISA-volume with the maximum-depth might be helpful to consider the endpoint of the PCI p...