Pulse Wave Velocity Prediction and Compliance Assessment in Elastic Arterial Segments

Abstract Pressure wave velocity (PWV) is commonly used as a clinical marker of vascular elasticity. Recent studies have increased clinical interest in also analyzing the impact of heart rate, blood pressure, and left ventricular ejection time on PWV. In this article we focus on the development of a theoretical one-dimensional model and validation via direct measurement of the impact of ejection time and peak pressure on PWV using an in vitro hemodynamic simulator. A simple nonlinear traveling wave model was developed for a compliant thin-walled elastic tube filled with an incompressible fluid. This model accounts for the convective fluid phenomena, elastic vessel deformation, radial motion, and inertia of the wall. An exact analytical solution for PWV is presented which incorporates peak pressure, ejection time, ejection volume, and modulus of elasticity. To assess arterial compliance, the solution is introduced in an alternative form, explicitly determining compliance of the wall as a function of the other variables. The model predicts PWV in good agreement with the measured values with a maximum difference of 3.0%. The results indicate an inverse quadratic relationship ( \(R^{2} = .99\) ) between ejection time and PWV, with ejection time dominating the PWV shifts (12%) over those observed with changes in peak pressure (2%). Our modeling and validation results both expla...
Source: Cardiovascular Engineering and Technology - Category: Cardiology Source Type: research