Study of Sensitivity and Resolution for Full Ring PET Prototypes based on Continuous Crystals and analytical modeling of the light distribution.
Study of Sensitivity and Resolution for Full Ring PET Prototypes based on Continuous Crystals and analytical modeling of the light distribution.
Phys Med Biol. 2018 Dec 11;:
Authors: Etxebeste A, Barrio J, Bernabéu Alberola J, Lacasta C, Llosa G, Muñoz E, Ros A, Oliver JF
Abstract
Sensitivity and spatial resolution are the main parameters to maximize in the performance of a PET scanner. For this purpose, detectors consisting of a combination of continuous crystals optically coupled to segmented photodetectors have been employed. With the use of continuous crystals the sensitivity is increased with respect to the pixelated crystals. In addition, spatial resolution is no longer limited to the crystal size. The main drawback is the difficulty in determining the interaction position. In this work, we present the characterization of the performance of a full ring based on cuboid continuous crystals coupled to SiPMs. To this end, we have employed the simulations developed in a previous work for our experimental detector head. Sensitivity could be further enhanced by using tapered crystals. This enhancement is obtained by increasing the solid angle coverage, reducing the wedge-shaped gaps between contiguous detectors. The performance of the scanners based on both crystal geometries was characterized following NEMA NU...