Demonstration of disinfection procedure for the development of accurate blood glucose meters in accordance with ISO 15197:2013

In this study, an effective disinfection procedure, conducting wiping 10 times to assure a one minute contact time of the disinfectant on contaminated surface, was first demonstrated using test samples of the meter housing materia ls, including acrylonitrile butadiene styrene (ABS), polymethyl methacrylate (PMMA), and polycarbonate (PC), in accordance with ISO 15197:2013. After bench studies comprising 10,000 disinfection cycles, the elemental compositions of the disinfected ABS, PMMA, and PC samples were almost the same as i n the original samples, as indicated by electron spectroscopy for chemical analysis. Subsequently, the validated disinfection procedure was then directly applied to disinfect 5 commercial BGM systems composed of ABS, PMMA, or PC to observe the effect of the validated disinfection procedure on meter accuracy. The results of HBsAg values after treatment with HBV sera and disinfectant wipes for each material were less than the LoD of each material of 0.020 IU/mL. Before and after the multiple disinfection cycles, 900 of 900 samples (100%) were within the system accuracy requirements of ISO 15197: 2013. All of the systems showed high performance before and after the series of disinfection cycles and met the ISO 15197:2013 requirements. In addition, our results demonstrated multiple cleaning and disinfection cycles that represented normal use over the lifetime of a meter of 3–5 years. Our va lidated cleaning and disinfection procedure can be directly applie...