EPID based in vivo dosimetry system: clinical experience and results.

EPID based in vivo dosimetry system: clinical experience and results. J Appl Clin Med Phys. 2016;17(3):6070 Authors: Celi S, Costa E, Wessels C, Mazal A, Fourquet A, Francois P Abstract Mandatory in several countries, in vivo dosimetry has been recognized as one of the next milestones in radiation oncology. Our department has implemented clinically an EPID based in vivo dosimetry system, EPIgray, by DOSISOFT S.A., since 2006. An analysis of the measurements per linac and energy over a two-year period was performed, which included a more detailed examination per technique and treat-ment site over a six-month period. A comparison of the treatment planning system doses and the doses estimated by EPIgray shows a mean of the differences of 1.9% (± 5.2%) for the two-year period. The 3D conformal treatment plans had a mean dose difference of 2.0% (± 4.9%), while for intensity-modulated radiotherapy and volumetric-modulated arc therapy treatments the mean dose difference was -3.0 (± 5.3%) and -2.5 (± 5.2%), respectively. In addition, root cause analyses were conducted on the in vivo dosimetry measurements of two breast cancer treatment techniques, as well as prostate treatments with intensity-modulated radiotherapy and volumetric-modulated arc therapy. During the breast study, the dose differences of breast treatments in supine position were correlated to patient setup and EPID positioning errors. Based on these observations, an automati...