Hormone anti-m üllérienne (AMH) en pédiatrie : établissement de valeurs de référence avec le dosage de l'AMH Fujirebio < sup > ® < /sup > Lumipulse G et comparaison avec le dosage Roche < sup > ® < /sup > Elecsys sur les m êmes échantillons pédiatriques

We described full pediatric gender and age-specific reference ranges for AMH using the Fujirebio® AMH assay on the Lumipulse G 600 II. The study was performed on 281 plasma samples collected in tubes with lithium heparin. The samples were from patients (135 males and 146 females) aged from 3 days to 22 years collected at the University Hospital Center of Tours. The Fujirebio® Lumipulse method showed excellent correlation with Roche® Elecsys but had a significant proportional positive bias. The data were used to propose pediatric reference values adapted to the Fujirebio® method. Our study described full pediatric gender and age-related reference ranges for AMH using the Fujirebio® AMH assay on the Lumipulse G600II. The delineation between normal male and female AMH concentrations make them valuable clinical tools for the monitoring of pediatric sexual and reproductive development from early childhood through the pubertal transition into adulthood.PMID:38638019 | DOI:10.1684/abc.2024.1867
Source: Annales de Biologie Clinique - Category: Biochemistry Authors: Source Type: research