Infections Associated with Medtronic Duet External Ventricular Drains - Rhode Island Hospital, Providence, Rhode Island, January 2023-January 2024

MMWR Morb Mortal Wkly Rep. 2024 Apr 11;73(14):312-316. doi: 10.15585/mmwr.mm7314a4.ABSTRACTExternal ventricular drains (EVDs) are medical devices that are inserted into the ventricles of the brain to drain excess fluid, manage intracranial hypertension, monitor intracranial pressure, and administer medications. Unintentional disconnections and breaks or fractures (breaks) of EVDs or associated drainage system components can result in cerebrospinal fluid (CSF) leakage and increased risk for EVD-associated infections. After replacement of Integra Life Sciences EVD systems with Medtronic Duet EVD systems at Rhode Island Hospital in mid-September 2023, a threefold increase was observed in the prevalence of positive CSF cultures, from 2.8 per 1,000 days with an EVD in place (EVD days) during January-September 2023 to 11.4 per 1,000 EVD days during October 2023-January 2024 (rate ratio [RR] = 5.7; 95% CI = 1.5-22.0; p = 0.01) and an eightfold increase in the prevalence of infections, from 0.7 to 6.5 per 1,000 EVD days (RR = 9.8; 95% CI = 1.1-87.3; p = 0.04). An investigation by Rhode Island Hospital Infection Control during December 2023-January 2024 identified frequent reports of disconnections and breaks of the Medtronic Duet EVD system. A search of the Food and Drug Administration Manufacturer and User Facility Device Experience database identified 326 reports nationwide of disconnection and breaks of components of the Duet EVD system, including 175 during 2023. A Medical Produc...
Source: MMWR Morb Mortal Wkl... - Category: Epidemiology Authors: Source Type: research