Effects of reverse osmosis membrane replacement of pure water system on clinical chemistry and immunoassay in clinical laboratory

CONCLUSIONS: Replacement of RO membrane could cause significant interference to TG test while have no effects on other laboratory tests performed in the study, which may be due to glycerol contamination. Our data provides important reference for management of test interference caused by RO membrane replacement. Clinical laboratory should observe the effects of RO membrane replacement on laboratory tests through both water quality monitoring and QC detection.PMID:38361738 | PMC:PMC10864026 | DOI:10.11613/BM.2024.010705
Source: Biochemia Medica - Category: Biochemistry Authors: Source Type: research