A high-throughput analytical method for complex contaminant mixtures in biosolids

In this study, a simplified, rapid, and robust method of analysis was developed and validated for a high-priority organic contaminant mixture of 44 endocrine disrupting compounds known to occur in biosolids. Analytes consisted of chemicals from many classes with a wide range of physiochemical properties (e.g., log Kow values from -1.4 to 8.9). The biosolids extraction and cleanup protocol was validated for 42 of the targeted compounds. The UPLC-MS2 parameters were validated for all 44 organic contaminants targeted for study. From the two batches of biosolids tested using this analytical method, most of the targeted contaminants (86%) were detected with 100% frequency at concentrations ranging from 0.036 to 10,226 μg/kg dw. Performance results highlighted that internal standards alone could not negate biosolids matrix effects; thus, internal standards and the standard addition method were used for residue quantification. This was the first study to detect and quantify 6PPD-q in biosolids, and the first to quantify lidocaine and 11 other chemicals in biosolids using a single analytical method. This method may be expanded for analysis of additional chemicals in biosolids and comparable matrices.PMID:38346636 | DOI:10.1016/j.envpol.2024.123517