Evaluation of correlation between Diopsys ® NOVA™ fixed-luminance flicker ERG and Diagnosys® Espion 2™ flicker ERG parameters

ConclusionsThere is a statistically significant positive correlation between light-adapted Diopsys ® NOVA™ fixed-luminance flickeramplitude and Diagnosys® flicker magnitude values. Additionally, there is a statistically significant positive correlation between Diopsys ® NOVA™ fixed-luminance flickerimplicit time (converted fromphase) and Diagnosys ® flicker implicit time values. These results imply that the Diopsys ® NOVA™ module, which utilizes the nonstandard shortened International Society for Clinical Electrophysiology of Vision (ISCEV) ERG protocol, can produce reliable light-adapted flicker ffERG measurements.
Source: Documenta Ophthalmologica - Category: Opthalmology Source Type: research