Individualized Scan Protocols in Abdominal Computed Tomography: Radiation Versus Contrast Media Dose Optimization

Background In contrast-enhanced abdominal computed tomography (CT), radiation and contrast media (CM) injection protocols are closely linked to each other, and therefore a combination is the basis for achieving optimal image quality. However, most studies focus on optimizing one or the other parameter separately. Purpose Reducing radiation dose may be most important for a young patient or a population in need of repetitive scanning, whereas CM reduction might be key in a population with insufficient renal function. The recently introduced technical solution, in the form of an automated tube voltage selection (ATVS) slider, might be helpful in this respect. The aim of the current study was to systematically evaluate feasibility of optimizing either radiation or CM dose in abdominal imaging compared with a combined approach. Methods Six Göttingen minipigs (mean weight, 38.9 ± 4.8 kg) were scanned on a third-generation dual-source CT. Automated tube voltage selection and automated tube current modulation techniques were used, with quality reference values of 120 kVref and 210 mAsref. Automated tube voltage selection was set at 90 kV semimode. Three different abdominal scan and CM protocols were compared intraindividually: (1) the standard “combined” protocol, with the ATVS slider position set at 7 and a body weight–adapted CM injection protocol of 350 mg I/kg body weight, iodine delivery rate (IDR) of 1.1 g I/s; (2) the CM dose-saving protocol, with the...
Source: Investigative Radiology - Category: Radiology Tags: Original Articles Source Type: research