Sequestration of Dexmedetomidine in Ex Vivo Cardiopulmonary Bypass Circuits

This study aimed to evaluate the disposition of DEX after administration to the ex vivo CPB circuits following different bolus doses and continuous infusion of DEX, including the effect of circuit coating, temperature, and modified ultrafiltration (MUF). Cardiopulmonary bypass circuits were setup ex vivo and primed with reconstituted blood. Dexmedetomidine was administered to the circuit (as a single bolus or single bolus along with continuous infusion). The circuit was allowed to equilibrate during the first 5 minutes, blood samples were collected at multiple time points (5–240 minutes). Blood samples were processed to collect plasma and analyzed for DEX with a validated assay. The majority of DEX sequestration in ex vivo CPB circuits occurred within the first 15 minutes. The percent of DEX remained in plasma pre-MUF (16–71%) and post-MUF (22–92%) varied depending on the dose and dosing scheme. Modified ultrafiltration significantly increased the plasma concentration of DEX in 19 of 23 circuits by an average of 12.1 ± 4.25% (p
Source: ASAIO Journal - Category: Medical Devices Tags: Pediatric Circulatory Support Source Type: research