Dosimetric comparison and validation of Eclipse Anisotropic Analytical Algorithm (AAA) and AcurosXB (AXB) algorithms in RapidArc-based radiosurgery plans of patients with solitary brain metastasis

This study aims to compare and validate Anisotropic Analytical Algorithm (AAA) and AcurosXB (AXB) algorithms of Eclipse Treatment Planning System (TPS) in RapidArc-based SRS plans of patients with solitary brain metastasis. Twenty patients with solitary brain metastasis who have been already treated with RapidArc SRS plans calculated using AAA plans were selected for this study. These plans were recalculated using AXB algorithm keeping the same arc orientations, multi-leaf collimator apertures, and monitor units.
Source: Medical Dosimetry - Category: Radiology Authors: Tags: Medical Physics Contribution Source Type: research
More News: Brain | Men | Neurology | Physics | Radiology | Study