Algorithm for an automatic treatment planning system using a single-arc VMAT for prostate cancer

J Appl Clin Med Phys. 2021 Oct 8. doi: 10.1002/acm2.13442. Online ahead of print.ABSTRACTOptimization process in treatment planning for intensity-modulated radiation therapy varies with the treatment planner. Therefore, a large variation in the quality of dose distribution is usually observed. To reduce variation, an automatic optimizing toolkit was developed for the Monaco treatment planning system (Elekta AB, Stockholm, Sweden) for prostate cancer using volumetric-modulated arc therapy (VMAT). This toolkit was able to create plans automatically. However, most plans needed two arcs per treatment to ensure the dose coverage for targets. For prostate cancer, providing a plan with a single arc was advisable in clinical practice because intrafraction motion management must be considered to irradiate accurately. The purpose of this work was to develop an automatic treatment planning system with a single arc per treatment for prostate cancer using VMAT. We designed the new algorithm for the automatic treatment planning system to use one arc per treatment for prostate cancer in Monaco. We constructed the system in two main steps: (1) Determine suitable cost function parameters for each case before optimization, and (2) repeat the calculation and optimization until the conditions for dose indices are fulfilled. To evaluate clinical suitability, the plan quality between manual planning and the automatic planning system was compared. Our system created the plans automatically in all p...
Source: Clinical Prostate Cancer - Category: Cancer & Oncology Authors: Source Type: research