Heparin-Free Extracorporeal Life Support Using Tethered Liquid Perfluorocarbon: A Feasibility and Efficacy Study

Coagulation management is the leading challenge during extracorporeal life support (ECLS) due to shear stress and foreign-surface–induced coagulation disturbance during circulation. A nonadhesive, liquid-infused coating called tethered liquid perfluorocarbon (TLP) was developed to prevent adhesion of blood on medical materials. We investigated the novel application of TLP to commercial ECLS circuits compared with standard heparin-coated circuits in vivo in anesthetized swine for 6 hours veno-venous ECLS (1 L/min blood flow) without systemic anticoagulation (n = 3/group). We hypothesized that TLP coating permits heparin-free circulation without untoward effects while reducing thrombus deposition compared with controls. Vital signs, respiration, gas transfer, coagulation, and histology were assessed. Scanning electron microscopy (SEM), elemental mapping, and digital imaging were used to assess thrombus deposition after circulation. There were no group differences in vitals, gas exchange, coagulation, and histology. In both groups, ECLS enabled a decrease in minute volume and end-tidal CO2, with concomitant increase in pH (p
Source: ASAIO Journal - Category: Medical Devices Tags: Pulmonary Source Type: research