Independent calculation-based verification of volumetric-modulated arc therapy-stereotactic body radiotherapy plans for lung cancer.

This study aimed to investigate the feasibility of independent calculation-based verification of volumetric-modulated arc therapy (VMAT)-stereotactic body radiotherapy (SBRT) for patients with lung cancer using a secondary treatment planning system (sTPS). In all, 50 patients with lung cancer who underwent VMAT-SBRT between April 2018 and May 2019 were included in this study. VMAT-SBRT plans were devised using the Collapsed-Cone Convolution in RayStation (primary TPS: pTPS). DICOM files were transferred to Eclipse software (sTPS), which utilized the Eclipse software, and the dose distribution was then recalculated using Acuros XB. For the verification of dose distribution in homogeneous phantoms, the differences among pTPS, sTPS, and measurements were evaluated using passing rates of a dose difference of 5% (DD5%) and gamma index of 3%/2 mm (γ3%/2 mm). The ArcCHECK cylindrical diode array was used for measurements. For independent verification of dose-volume parameters per the patient's geometry, dose-volume indices for the planning target volume (PTV) including D95% and the isocenter dose were evaluated. The mean differences (± standard deviations) between the pTPS and sTPS were then calculated. The gamma passing rates of DD5% and γ3%/2 mm criteria were 99.2 ± 2.4% and 98.6 ± 3.2% for pTPS vs. sTPS, 92.9 ± 4.0% and 94.1 ± 3.3% for pTPS vs. measurement, and 93.0 ± 4.4% and 94.3 ± 4.1% for sTPS vs. measurement, respectively. The differences between pTPS...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research