Platelet-Derived Microparticles Generated by Neonatal Extracorporeal Membrane Oxygenation Systems

The objective of this study was to compare PMP generation in five different neonatal ECMO systems, using a simulated circuit with swine blood at 300 ml/min for 4 hours. Systems were composed of both newer components (centrifugal pump and hollow-fiber oxygenator) and traditional components (roller-head pump and silicone membrane oxygenator). Free plasma hemoglobin levels were measured as an indicator of hemolysis and flow cytometry-measured PMP. Hemolysis generated in all ECMO systems was similar to that observed in noncirculated static blood (p = 0.48). There was no difference in net PMP levels between different oxygenators with a given pump. In contrast, net PMP generation in ECMO systems with a centrifugal pump was at least 2.5 times greater than in roller-head pump systems. This was significant when using either a hollow-fiber (p < 0.005) or a silicone membrane (p < 0.05) oxygenator. Future studies are needed to define the relationship between pump-generated PMP and thrombosis.
Source: ASAIO Journal - Category: Medical Equipment Tags: Pediatric Circulatory Support Source Type: research