Influence of photo-initiators in the preparation of methacrylate monoliths into poly(ethylene-co-tetrafluoroethylene) tubing for microbore HPLC.

In this study, poly(butyl methacrylate-co-ethyleneglycol dimethacrylate) polymeric monoliths were in situ developed within 0.75 mm i.d. poly(ethylene-co-tetrafluoroethylene) (ETFE) tubing by UV polymerization via three different free-radical initiators (α,α'-azobisisobutyronitrile (AIBN), 2,2-dimethoxy-2-phenylacetophenone (DMPA) and 2-methyl-4'-(methylthio)-2-morpholinopropiophenone (MTMPP). The influence of the nature of each photo-initiator and irradiation time on the morphological features of the polymer was investigated by scanning electron microscopy, and the chromatographic properties of the resulting microbore columns were evaluated using alkyl benzenes as test substances. The beds photo-initiated with MTMPP gave the best performance (minimum plate heights of 38 μm for alkyl benzenes) and exhibited a satisfactory reproducibility in the chromatographic parameters (RSD < 11%). These monolithic columns were also successfully applied to the separation of phenylurea herbicides, proteins and a tryptic digest of β-casein. PMID: 31735210 [PubMed - in process]
Source: Analytica Chimica Acta - Category: Chemistry Authors: Tags: Anal Chim Acta Source Type: research