Evaluation of scattered radiation from fluoroscopy using small OSL dosimeters

In this study, a small optically stimulated luminescence (OSL) dosimeter called the nanoDot was used to evaluate and measure the amount of radiation; this dosimeter was developed by Landauer Ltd. and was specifically designed for point measurements. There are numerous studies that have reported the usefulness of personal OSL dosimeters other than the nanoDot to measure scattered radiation. Here, we evaluated the amount of scattered radiation, along with the degree of reduction achieved with the use of a protective curtain, while employing a personal OSL dosimeter and nanoDot. When the scattered radiation dose was measured using the nanoDot, the maximum recorded value without a protective curtain was 0.363  mGy and that with a protective curtain was 0.026 mGy, both at the height of 100 cm. The maximum reduction rate of scattered radiation while using a protective curtain was approximately 93% and 97% at 100 cm and 150 cm, respectively. The measured values recorded using both personal OSL dosimete rs and nanoDot machine were strongly correlated.
Source: Radiological Physics and Technology - Category: Physics Source Type: research
More News: ERCP | Nanotechnology | Physics | Study