Case of an In-Laboratory Vagal Nerve Stimulator Titration for Vagal Nerve Stimulator-Induced Pediatric Obstructive Sleep Apnea

We present the case of a 12-year-old girl with medically refractory epilepsy and a vagal nerve stimulator (VNS), who experienced severe obstructive sleep apnea (OSA) with respiratory events closely matching her VNS settings. We demonstrated a real-time decrease in OSA through an in-laboratory VNS titration study, decreasing her VNS frequency from 20 Hz to 10 Hz. We were able to demonstrate a baseline level of OSA by turning off the VNS. We then effectively treated her residual OSA with continuous positive airway pressure (CPAP). Novel to our case is that this in-laboratory VNS titration did not result in any subsequent increase in seizure frequency. After 5 months, her seizure frequency had decreased. Our case demonstrates that in-laboratory VNS titration can be an efficient tool for optimizing treatment of VNS-induced OSA and assert that polysomnography before VNS placement is important for guiding future care.Citation:Chan J, Owens JW, Wrede JE. Case of an in-laboratory vagal nerve stimulator titration for vagal nerve stimulator-induced pediatric obstructive sleep apnea.J Clin Sleep Med. 2019;15(10):1539–1542.
Source: Journal of Clinical Sleep Medicine : JCSM - Category: Sleep Medicine Source Type: research