Sensors, Vol. 19, Pages 3706: Development of Open-Tubular-Type Micro Gas Chromatography Column with Bump Structures

This study reports an open-tubular-type micro GC (μ-GC) column with internal bump structures (bump structure μ-GC column) that substantially increase the interaction between the gas mixture and a stationary phase. The developed bump structure μ-GC column, which was fabricated on a 2 cm × 2 cm μ-GC chip and coated with a non-polar stationary phase, is 1.5 m-long, 150 μm-wide, and 400 μm-deep. It has an internal microfluidic channel in which the bumps, which are 150 μm diameter half-circles, are alternatingly disposed to face each other on the surface of the microchannel. The fabricated bump structure μ-GC column yielded a height-equivalent-to-a-theoretical-plate (HETP) of 0.009 cm (11,110 plates/m) at an optimal carrier gas velocity of 17 cm/s. The mechanically robust bump structure μ-GC column proposed in this study achieved higher separation efficiency than a commercially available GC column and a typical μ-GC column with internal post structures classified as a semi-packed-type column. The experimental results demonstrate that the developed bump structure μ-GC column can separate a gas mixture completely, with excellent separation resolution for formaldehyde, benzene, toluene, ethylbenzene, and xylene mixture, under programmed operating temperatures.
Source: Sensors - Category: Biotechnology Authors: Tags: Article Source Type: research