Investigating the Accuracy of Ultrasound-Ultrasound Fusion Imaging for Evaluating the Ablation Effect via Special Phantom-Simulated Liver Tumors

The goal of this study was to investigate the accuracy of ultrasound-ultrasound (US-US) fusion imaging for evaluating the ablation effect via phantom-simulated liver tumors. Twenty special phantom models were established, ablated and divided into a complete ablation group (n  = 10) and an incomplete ablation group (n = 10). US-US fusion imaging was performed to evaluate the ablation effect. Gross specimens were observed as a standard reference. In this US-US fusion imaging study, the registration success rate was 100% (20/20), and the assessment time was 3.8 ± 0.9 min.