Methodological Considerations for a Vascular Function Test Battery

This study aimed to determine the test-retest reliability of a test battery of vascular function measures: automated blood pressure (BP), laser Doppler imaging with iontophoresis (LDI), digital volume pulse (DVP), pulse wave velocity (PWV), augmentation index (AIx) measured by pulse wave analysis (PWA) and flow-mediated dilation (FMD) taken within and between sessions. Measures were taken in 21 non-smoking males intra-session and again inter-session (one week apart) to determine repeatability and reproducibility, respectively. There was moderate to excellent repeatability (ICC: 0.53–0.93; CV=2.2–18.1%) and reproducibility (ICC: 0.71–0.96; CV 1.9–14.2%) for BP, DVP stiffness index, PWV, AIx, AIx normalised to heart rate (75 bpm), absolute and percentage FMD. Repeatability of the DVP reflection index was moderate (ICC: 0.64; CV=9.5%) but there was poor reproducibility (ICC: 0.17; CV=15.1%). Moreover, the repeatability and reproducibility of the LDI measures ranged from poor to good (ICC: 0.31–0.84; CV=28.4–36.7%). These data indicated that there was considerable variability in the repeatability and reproducibility of measurements of endothelial function and arterial stiffness taken in a battery of measurements, which needs careful consideration in future research designs. [...] © Georg Thieme Verlag KG Stuttgart · New YorkArticle in Thieme eJournals: Table of contents  |  Abstract  |  Full text