Accuracy of computed tomography-magnetic resonance imaging image fusion using a phantom for skull base surgery.

CONCLUSIONS: The displacements following image fusion were within approximately 0.5 mm, even at 8 cm from the center. This suggests that the region where the cranial nerves are located, within 40 mm of the dorsum sellae, had no risk of positional error following image fusion. PMID: 30808859 [PubMed - as supplied by publisher]
Source: Journal of Neurosurgical Sciences - Category: Neurosurgery Tags: J Neurosurg Sci Source Type: research