IMRT dose verification considering passing rate and respiratory motion.

IMRT dose verification considering passing rate and respiratory motion. Oncol Lett. 2018 Jul;16(1):963-969 Authors: Xie K, Sun H, Lin T, Gao L, Sui J, Ni X Abstract The aim of the present study was to investigate the association between the dynamic intensity-modulated radiation therapy planned γ analysis passing rate and respiratory amplitude (A) and period (T) for different tumor volumes. A total of 30 patients with malignant lung tumors were divided into three groups: A; B; and C. The average tumor volumes (V) in the A, B and C groups were 635, 402 and 213 cm3, respectively. The simulated A values were set at 0, 5, 10, 15, 20 and 25 mm. The T values were set at 4, 5 and 6 sec. The γ analysis passing rate was calculated under different conditions (dose difference, 3%; distance difference, 3 mm). Compared with the γ analysis passing rate in the A group (A=0, static; T=4, 5, 6 sec), the γ analysis passing rate deviation (A=5 mm) was <3.3%. However, this difference was not statistically significant (P>0.05). With a gradual increase in A value, the passing rate decreased. The deviation between the 3 groups was <2.5% at the same A value (T=4, 5 and 6 sec). A descending trend of passing rate with increased A value was revealed. At the same A and T values, the passing rate decreased with decreased tumor volume. At the same tumor volume, the passing rate decreased when the A value increased. The respiratory cycle was not demons...
Source: Oncology Letters - Category: Cancer & Oncology Tags: Oncol Lett Source Type: research