Dosimetric validation of a magnetic resonance image gated radiotherapy system using a motion phantom and radiochromic film.

CONCLUSIONS: Our results suggest that the device can be used to compensate respiratory motion using a 3 mm gating margin and 10% allowed excursion results in conjunction with repeated breath holds. Full clinical validation requires a comprehensive evaluation of tracking performance in actual patient images, outside the scope of this study. PMID: 28436094 [PubMed - as supplied by publisher]
Source: Journal of Applied Clinical Medical Physics - Category: Physics Authors: Tags: J Appl Clin Med Phys Source Type: research