Utilization of a Novel, Multi-Durometer Intracranial Distal Access Catheter: Nuances and Experience in 110 Consecutive Cases of Aneurysm Flow Diversion

We report the first experience using Cat5 for aneurysm treatment by flow diversion.Methods: A single-center aneurysm database was reviewed for cases of aneurysm treatment with the Pipeline embolization device (PED) that utilized Cat5. Data were collected for patient demographics, aneurysm characteristics, procedural details, catheter positions, vessel tortuosity, and catheter related complications.Results: One hundred and ten cases of aneurysm flow diversion were successfully performed using Cat5. Patient age ranged from 21 to 86 years (mean 57 ± 12.5 years) with 84% women. Aneurysm size ranged from 2 to 28 mm (mean 5.7 ± 5.0 mm), with 97% in the anterior circulation. Twenty-four aneurysms (22%) were located beyond the ICA termination. Significant cervical carotid tortuosity was present in 26% of cases, and moderate to severe cavernous t ortuosity (cavernous grade ≥2) in 45% of cases. Cat5 was tracked to the intended distal position in all cases with 100% technical success of PED implantation. No iatrogenic catheter-related vessel injury occurred, and major neurological morbidity occurred in 1 patient (1%).Summary: The Cat5 is a novel, multi-durometer cranial distal access catheter designed for use in tri-axial systems. We have demonstrated the utility of Cat5 in 110 successful cases of flow diversion with a wide range of complexity. This catheter is a new tool in the neurointerventionalist's armamentarium to achieve robust and atraumatic distal access.Intervent Neurol 20...