A comparative study based on image quality and clinical task performance for CT reconstruction algorithms in radiotherapy

In this study we designed and implemented a new comparative analysis method associating image quality, radiation dose, and patient size with radiotherapy task performance, with the purpose of guiding the clinical radiotherapy usage of CT reconstruction algorithms. The iDose4 iterative reconstruction algorithm was selected as the target for comparison, wherein filtered back-projection (FBP) reconstruction was regarded as the baseline. Both phantom and patient images were analyzed. A layer-adjustable anthropomorphic pelvis phantom capable of mimicking 38–58 cm lateral diameter-sized patients was imaged and reconstructed by the FBP and iDose4 algorithms with varying noise-reduction-levels, respectively. The resulting image sets were quantitatively assessed by two image quality indices, noise and contrast-to-noise ratio, and two clinical task-based indices, target CT Hounsfield number (for electron density determination) and structure contouring accuracy (for dose-volume calculations). Additionally, CT images of 34 patients reconstructed with iDose4 with six noise reduction levels were qualitatively evaluated by two radiation oncologists using a five-point scoring mechanism. For the phantom experiments, iDose4 achieved noise reduction up to 66.1% and CNR improvement up to 53.2%, compared to FBP without considering the changes of spatial resolution among images and the clinical acceptance of reconstructed images. Such improvements consistently appeared across different iDose4 no...
Source: Journal of Applied Clinical Medical Physics - Category: Physics Source Type: research