Artificial intelligence-enhanced automation for M-mode echocardiographic analysis: ensuring fully automated, reliable, and reproducible measurements

AbstractTo enhance M-mode echocardiography ’s utility for measuring cardiac structures, we developed and evaluated an artificial intelligence (AI)-based automated analysis system for M-mode images through the aorta and left atrium [M-mode (Ao-LA)], and through the left ventricle [M-mode (LV)]. Our system, integrating two deep neural networ ks (DNN) for view classification and image segmentation, alongside an auto-measurement algorithm, was developed using 5,958 M-mode images [3,258 M-mode (LA-Ao), and 2,700 M-mode (LV)] drawn from a nationwide echocardiographic dataset collated from five tertiary hospitals. The performance of view classification and segmentation DNNs were evaluated on 594 M-mode images, while automatic measurement accuracy was tested on separate internal test set with 100 M-mode images as well as external test set with 280 images (140 sinus rhythm and 140 atrial fibrillation). Performance evaluation showed the view classification DNN’s overall accuracy of 99.8% and segmentation DNN’s Dice similarity coefficient of 94.3%. Within the internal test set, all automated measurements, including LA, Ao, and LV wall and cavity, resonated strongly with expert evaluations, exhibiting Pearson’s correlation coefficients (PCCs) of 0.81–0.99. This performance persisted in the external test set for both sinus rhythm (PCC, 0.84–0.98) and atrial fibrillation (PCC, 0.70–0.97). Notably, automatic measurements, consistently offering multi-cardiac cycle readings...
Source: The International Journal of Cardiovascular Imaging - Category: Radiology Source Type: research