Cardiac implantable electronic device ‐induced tricuspid regurgitation: Implications and management

AbstractTricuspid regurgitation (TR) secondary to cardiac implantable electronic devices (CIEDs) has been well documented and is associated with worse cardiovascular outcomes. A variety of mechanisms have been proposed including lead-induced mechanical disruption of the tricuspid valvular or subvalvular apparatus and pacing-induced electrical dyssynchrony. Patient characteristics such as age, sex, baseline atrial fibrillation, and pre-existing TR have not been consistent predictors of CIED-induced TR. While two-dimensional echocardiography is helpful in assessing the severity of TR, three-dimensional echocardiography has significantly improved accuracy in identifying the etiology of TR and whether lead position contributes to TR. Three-dimensional echocardiography may therefore play a future role in optimizing lead positioning during implant to reduce the risk of CIED-induced TR. Optimal lead management strategies in addition to percutaneous interventions and surgery in alleviating TR are very important.
Source: Journal of Cardiovascular Electrophysiology - Category: Cardiology Authors: Tags: REVIEW Source Type: research